iVario Breakout

iVario Breakout
PN OE-565-BASE
The iVario Breakout is a discrete kit meant to make the set up and installation of any COMET iXRS series x-ray source simple and easy. The unit brings all iVario connections to labeled terminal blocks, getting rid of the need for customized connectors/cables.
The iVario Breakout includes interfaces for:
-
Generator mains and auxiliary power
-
Cooler/chiller power
-
High voltage door interlock (CDRH)
-
Three monitored warning lamps
-
Four software controlled potential-free outputs
-
Safety interlocks, including: emergency stops, low voltage door interlocks, etc.
New Product Version Features (shipping 11/2020, SN565136):
-
OUTPUT3 and OUTPUT4
-
Now rated for 250VAC/220VDC @ 2A
-
Now fused @ 2A with replaceable on-board fuses
-
**NOTE** OUTPUT1 and OUTPUT2 are still rated for 50VDC @ 0.5A
-
-
New Mounting Options:
-
PCB can now mount directly to 35mm DIN rail with included hardware
-
PCB can now mount directly to a panel or wall
-
-
Additional 24VDC terminal blocks added
-
Terminal blocks now labeled "+" and "-" for 24VDC and common, respectively
-
-
Additional earth ground terminal blocks added
-
Terminal blocks now labeled "⏚"
-
For complete details, please see the Product Change Notice

Additional Details:
-
All power outputs fused with branch circuit Class CC fuses [UL/CSA/CE]
-
iVario Mains - 15A x4
-
iVario Auxiliary - 3A x2
-
Cooler/Chiller - 10A x2
-
Compliant with UL61010-1 / CSA C22.2 No. 61010-1 Third Edition, Safety requirements for electrical equipment for measurement, control and laboratory use - See MetLabs Special Report 100834